Objectives The accuracy of point-of-care blood glucometers in the detection and evaluation of neonatal hypoglycemia is a concern. This study compared the performance of three i-SENS glucometers with that of the YSI 2300 STAT Plus Analyzer, which was used as a reference. Methods The leftover neonatal capillary blood samples of 319 patients were used in this study. The evaluation process and accuracy performance criteria were based on the International Organization for Standardization 15197:2013 guidelines. The evaluation involved three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK ® Inform II glucometer. Results The accuracy evaluation yielded acceptable results as follows: a) 100 and 100% for BAROzen H Expert plus; 99.8 and 100% for CareSens PRO; 98.7%, and 97.2% for CareSens H Beat glucometers were within the range of ±0.8 mmol/L (15 mg/dL) and ±15% of the average reference values at glucose concentrations <5.55 mmol/L (100 mg/dL) and ≥5.55 mmol/L (100 mg/dL), respectively, and b) all estimated glucose values (100%) were within the zones A and B of Consensus Error Grid for all three i-SENS glucometers. There was good correlation between the glucose values estimated by the glucometers and the reference values (R>0.990). Conclusions This study demonstrated that i-SENS glucometers exhibit acceptable performance and can be used as effective point-of-care devices in neonates.