Key result
Aortic root restoration using the HAART 300 ring shows 0% mortality and major complications.
Why the study?
The study evaluated the safety and efficacy of the HAART 300 internal geometric annuloplasty ring in aortic valve repair.
Does the HAART 300 internal annuloplasty ring safely and effectively improve aortic valve regurgitation in patients with trileaflet AV insufficiency and aortic root enlargement?
Observational (n=20)
No
Does the HAART 300 internal annuloplasty ring safely and effectively improve aortic valve regurgitation in patients with trileaflet AV insufficiency and aortic root enlargement?
The HAART 300 internal annuloplasty ring is a safe and effective valve-sparing approach for treating aortic valve insufficiency associated with root enlargement, demonstrating excellent short-term clinical and echocardiographic outcomes.
Answers are grounded in this paper and cite their sources.
Supports early safety data for HAART 300; leaves open durability and generalizability pending larger studies.
BACKGROUND AND AIM OF THE STUDY: HAART 300 is an internal geometric annuloplasty ring. The safety and efficacy of this novel device in aortic valve (AV) repair in a single referral center are reported. METHODS: Twenty patients with trileaflet AV insufficiency with ascending aorta and/or aortic root enlargement were included. Subannular implantation was performed to correct annular dilatation, whereas concomitant leaflet repair was performed whenever required. All but two patients also received ascending aorta replacement, whereas selective sinus replacement was performed in all but five patients. RESULTS: Follow-up was for a maximum of 3.8 years and a mean of 2.2 years. Mean age was 54.2 years old. Moderate to severe preoperative AV insufficiency was noted in 75% of patients, whereas 70% of them had an ascending aorta over 45 mm. One patient was lost from follow-up. Overall mortality as well as major complication rates were zero. Early postoperatively, no more than mild AV regurgitation was detected, whereas only one patient appeared with moderate AV regurgitation during our 2.2-year follow-up. New York Heart Association class was also significantly lower compared to preoperative values and valve gradients remained low at last follow-up. CONCLUSIONS: Geometric ring annuloplasty is a safe and effective valve sparing approach to deal with AV insufficiency contributing to overall root reconstruction. Short-term results are excellent rendering this easily reproducible and versatile method very attractive.
No takes yet. Share an insight, caveat, or question.
Papakonstantinou et al. (2021) conducted an observational in Trileaflet aortic valve insufficiency (n=20). HAART 300 internal geometric annuloplasty ring was evaluated on Overall mortality and major complications. Aortic root restoration using the HAART 300 internal annuloplasty ring resulted in 0% overall mortality and major complications over a mean follow-up of 2.2 years.
Synapse has enriched 10 related papers. Consider them for comparative context: