Nabz Hooshmand-1 handheld ECG achieves 98% morphology matching vs standard GE systems to detect pathological rhythms.
Why the study?
The widespread use of diverse wireless electrocardiography devices warranted validation of their accuracy and reliability for remote cardiac monitoring and telemedicine.
Does the Nabz Hooshmand-1 handheld ECG device accurately measure ECG parameters compared to the GE MAC 1200 ECG system in inpatients?
Cross-Sectional (n=300)
Does the Nabz Hooshmand-1 handheld ECG device accurately measure ECG parameters compared to the GE MAC 1200 ECG system in inpatients?
The Nabz Hooshmand-1 handheld ECG device is highly accurate and reliable for rhythm detection and basic ECG parameter measurement compared to a standard 12-lead ECG system.
Answers are grounded in this paper and cite their sources.
Supports preliminary handheld ECG accuracy for rhythm detection; leaves open prospective validation before clinical adoption.
BACKGROUND: Remote cardiac monitoring and screening have already become an integral telemedicine component. The wide usage of several different wireless electrocardiography (ECG) devices warrants a validation study on their accuracy and reliability. METHODS: Totally, 300 inpatients with the Nabz Hooshmand-1 handheld ECG device and the GE MAC 1200 ECG system (as the reference) were studied to check the accuracy of the devices in 1 and 6-limb lead performance. Simultaneous 10-second resting ECGs were assessed for the most common ECG parameters in lead I. Afterward, 6-lead ECGs (limb leads), were performed immediately and studied for their morphologies. RESULTS: Of the 300 patients, 297 had acceptable ECG quality in both devices for simultaneous lead I ECGs. The ECGs were inspected on-screen by a cardiologist for their rhythms, rates, axes, numbers, morphologies of premature atrial and ventricular beats, morphologies and amplitudes of PQRST waves, P-wave durations, QRS-wave durations, P-R intervals, and QT intervals. No significant differences were detected between the devices, and no major abnormalities were missed. Six-limb lead ECGs were obtained in 284 patients, of whom 281 had acceptable quality in ECGs by both devices. The morphology matching evaluation of the ECGs demonstrated an overall 98% compatibility rate, with the highest compatibility in lead I and the lowest in lead augmented vector foot. CONCLUSIONS: The diagnosis of critical pathological rhythms, including atrial fibrillation and high-grade atrioventricular node block, was not missed by the Nabz Hooshmand-1 and GE MAC 1200 ECG devices. Accordingly, rhythm detection as the primary purpose of handheld ECG devices was highly accurate. Both devices had acceptable sensitivity to diagnose long P-R and long and short QT intervals. Although the modern technology of smartphones and the physical inability for the 6-limb mode might cause old patients difficulty in utilizing such devices, their use for screening and follow-up is safe.
No takes yet. Share an insight, caveat, or question.
Ahmadi‐Renani et al. (2022) conducted a cross-sectional in Inpatients requiring ECG (n=300). Nabz Hooshmand-1 handheld ECG device vs. GE MAC 1200 ECG system was evaluated on Accuracy and reliability in 1 and 6-limb lead performance (morphology matching compatibility). The Nabz Hooshmand-1 handheld ECG device demonstrated an overall 98% morphology matching compatibility rate compared to the GE MAC 1200 system, accurately detecting critical pathological rhythms.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: