Objective There is a debate if the [ NO ] required to influence vascular smooth muscle is below 50 nM or much higher. Electrodes with 30 μm and larger diameter report [ NO ] below 50 nM, whereas those with diameters of <10–12 μm report hundreds of nM. This study examined how size of electrodes influenced [ NO ] measurement due to NO consumption and unstirred layer issues. Methods Electrodes were 2 mm disk, 30 μm × 2 mm carbon fiber, and single 7 μm diameter carbon fiber within open tip microelectrode, and exposed 7 μm carbon fiber of ~15 μm to 2 mm length. Results All electrodes demonstrated linear calibrations with sufficient stirring. As stirring slowed, 30 μm and 2 mm electrodes reported much lower [ NO ] due to unstirred layers and high NO consumption. The three 7 μm microelectrodes had minor stirring issues. With limited stirring with NO present, 7 μm open tip microelectrodes advanced toward 30 μm and 2 mm electrodes experienced dramatically decreased current within 10–50 μm of the larger electrodes due to high NO consumption. None of the 7 μm microelectrodes interacted. Conclusions The data indicate large electrodes underestimate [ NO ] due to excessive NO consumption under conditions where unstirred layers are unavoidable and true microelectrodes are required for valid measurements.