Key result
New MRI-conditional pacemakers incorporate design changes such as reduced ferromagnetic components and modified leads to safely allow magnetic resonance imaging in patients with pacemakers.
Why the study?
Are magnetic resonance imaging-safe pacemakers safe for patients requiring MRI scans?
Are magnetic resonance imaging-safe pacemakers safe for patients requiring MRI scans?
Answers are grounded in this paper and cite their sources.
New MRI-conditional pacemakers incorporate design changes that mitigate the risks of MRI scanning, offering a safe imaging option for patients who require pacing.
Due of its superior soft tissue imaging capabilities, magnetic resonance imaging (MRI) has become the imaging modality of choice in many clinical situations, as illustrated by the tremendous growth in the number of MRIs performed over the past 2 decades. In parallel, the number of patients who require pacemakers or implantable cardiac defibrillators is increasing as indications for these devices broaden and the population ages. Taken together, these phenomena present an important clinical issue, as MR scans are generally contraindicated-except in urgent situations-in patients who have implanted cardiovascular devices. Potentially deleterious interactions between the magnetic fields and radio frequency (RF) energy produced by MR equipment and implantable devices have been identified, including inhibition of pacing, asynchronous/high-rate pacing, lead tip heating, and loss of capture. New devices that incorporate technologies to improve MR safety in patients with pacemakers have recently received approval in Europe and are under evaluation in the United States. Initial data from these devices suggest that these devices are safe in the MRI environment.
No takes yet. Share an insight, caveat, or question.
Jung et al. (2011) conducted a review in Patients with pacemakers or implantable cardiac defibrillators requiring MRI. MRI-safe pacemakers was evaluated. New MRI-conditional pacemakers incorporate design changes such as reduced ferromagnetic components and modified leads to safely allow magnetic resonance imaging in patients with pacemakers.
Synapse has enriched 10 related papers. Consider them for comparative context: