Objective: This study aimed to investigate the temperature changes and question the safe laser settings and protocols for laser-assisted peri-implantitis treatment in an in vitro environment. Materials and methods: Three types of implants (Neoss, Dentegris, and Camlog) were implanted in an artificial bone model (n = 15). The model was placed into a 37°C water bath to simulate the in vivo oral condition. Throughout the laser irradiation, K-type thermocouples were used to record the real-time temperature measurements at different anatomically important locations in the artificial bone. Results: In all of the temperature measurements, no temperature rise above the critical safe limit, that is, >47°C, was observed. Conclusions: Within the limitations of this in vitro study, the use of the dual-wavelength protocol [Er,Cr:YSGG (1 and 1.5 W, 25, 50, 75 Hz)] and the 940 nm diode (peak power 2 W, duty cycle 50%) can be considered a safe method in the treatment of peri-implantitis, regarding its thermal safety.