Key result
Novel RF wrist sensor detects heart rate across a 2 mm gap from the skin.
A novel nonconstrained wrist pulse sensor using a 2.4 GHz RF resonator and PLL can successfully detect heart rate at a 2 mm distance from the skin.
In this paper, a nonconstrained wrist pulse sensor based on radio-frequency proximity coupling at 2.4 GHz is proposed. The proposed system consists of a voltage-controlled oscillator (VCO) combined with a planar resonator and a phase-locked loop (PLL). The resonator senses the distance variation from the skin occurring because of the wrist pulse and plays a role as a feedback element of the VCO. The oscillation frequency of the VCO varies due to the impedance variation of the resonator, and the PLL operates as a converter of the frequency variation to the DC output voltage variation. The measurement result verified that the proposed system can detect the heart rate in nonconctact and nonconstrained condition with 2 mm gap from the skin.
No takes yet. Share an insight, caveat, or question.
Baek et al. (2021) studied Heart rate detection. Nonconstrained wrist pulse sensor based on RF proximity coupling at 2.4 GHz was evaluated on Heart rate detection in noncontact condition. A proposed nonconstrained wrist pulse sensor based on a 2.4 GHz RF resonator and PLL successfully detected heart rate in a noncontact condition with a 2 mm gap from the skin.