A post-processing technique has been developed to suppress the magnetic resonance imaging (MRI) artifact arising from object planar rigid motion. In two-dimensional Fourier transform (2-DFT) MRI, rotational and translational motions of the target during magnetic resonance magnetic resonance (MR) scan respectively impose nonuniform sampling and a phase error an the collected MRI signal. The artifact correction method introduced considers the following three conditions: (1) for planar rigid motion with known parameters, a reconstruction algorithm based on bilinear interpolation and the super-position method is employed to remove the MRI artifact, (2) for planar rigid motion with known rotation angle and unknown translational motion (including an unknown rotation center), first, a super-position bilinear interpolation algorithm is used to eliminate artifact due to rotation about the center of the imaging plane, following which a phase correction algorithm is applied to reduce the remaining phase error of the MRI signal, and (3) to estimate unknown parameters of a rigid motion, a minimum energy method is proposed which utilizes the fact that planar rigid motion increases the measured energy of an ideal MR image outside the boundary of the imaging object; by using this property all unknown parameters of a typical rigid motion are accurately estimated in the presence of noise. To confirm the feasibility of employing the proposed method in a clinical setting, the technique was used to reduce unknown rigid motion artifact arising from the head movements of two volunteers.
No takes yet. Share an insight, caveat, or question.
Zoroofi et al. (1996) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: