A novel mathematical model based on electromagnetic induction accurately predicts arterial stenosis rates in simulations, providing a theoretical basis for a new non-invasive diagnostic method.
Answers are grounded in this paper and cite their sources.
May support electromagnetic models for stenosis assessment; leaves open human validation before clinical consideration.
This paper presents a mathematical model of measuring blood flow based on electromagnetic induction for predicting the rate of arterial stenosis. Firstly, an electrode sensor was used to collect the induced potential differences from human skin surface in a uniform magnetic field. Then, the inversion matrix was constructed by the weight function theory and finite element method. Next, the blood flow volume inversion model was constructed by combining the induction potential differences and inversion matrix. Finally, the rate of arterial stenosis was predicted based on mathematical relationship between blood flow and the area of arterial stenosis. To verify the accuracy of the model, a uniform magnetic field distribution of Helmholtz coil and a 3D geometric model of the ulnar artery of the forearm with different rates of stenosis were established in COMSOL, a finite element analysis software. Simulation results showed that the inversion model had high accuracy in the measurement of blood flow and the prediction of rate of stenosis, and is of great significance for the early diagnosis of arterial stenosis and other vessel diseases.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2019) studied this question.