First-time pulmonary vein isolation using the variable-loop circular catheter achieved a 12-month primary effectiveness success of 74.8% and a primary adverse event rate of 1.6%.
Does pulsed field ablation using a variable-loop circular catheter improve 12-month freedom from atrial arrhythmia recurrence in Chinese patients with symptomatic, drug-refractory paroxysmal atrial fibrillation?
Pulsed field ablation using a variable-loop circular catheter demonstrated favorable 12-month effectiveness and safety for pulmonary vein isolation in a Chinese population with paroxysmal atrial fibrillation.
BACKGROUND: The variable-loop circular catheter (VLCC) showed promising outcomes in European and US studies, but evidence is limited in Asian populations. afIRE trial assessed 12-month safety and effectiveness of the VLCC in Chinese patients with symptomatic, drug-refractory paroxysmal atrial fibrillation (AF). METHODS: AfIRE was a prospective, multicenter, single-arm trial for adults undergoing first-time pulmonary vein isolation (PVI) with the VLCC. Primary effectiveness was 12-month freedom from atrial arrhythmia recurrence (systematically assessed by electrocardiogram and Holter monitoring at 3-, 6-, and 12-months post-ablation or when symptomatic), repeat ablation, procedural failure, or Class I/III antiarrhythmic drug use. Safety was defined by incidence of primary adverse events (PAEs). RESULTS: Of 142 patients enrolled, 123 (mean age 57.0 years, 45.5% female) comprised the main-stage cohort (patients treated with the VLCC after roll in). Median procedure, pulsed field ablation (PFA), and fluoroscopy times were 115.0, 36.1, and 5.0 min, respectively. Acute PVI was achieved in all patients, with 89.4% without acute reconnection and 11.4% undergoing non-PV ablation. Twelve-month primary effectiveness success and freedom from symptomatic recurrence were 74.8% and 87.0%, respectively. Safety was favorable, with PAEs in 2 patients (1.6%; vascular complications; unlikely device-related). Multivariable analysis of the primary endpoint showed that preserved left ventricular ejection fraction (≥ 60%) was associated with 86% reduction in odds of 12-month atrial arrhythmia recurrence; diabetes, ≥ 9 symptomatic AF episodes, and < 48 valid PFA applications (16 ablations) numerically increased odds of recurrence. CONCLUSION: The first Asian VLCC experience demonstrated favorable 12-month effectiveness and excellent safety, reinforcing reproducibility across global AF populations. CLINICAL TRIAL REGISTRATION: NCT05552963.
Tao et al. (Thu,) conducted a other in symptomatic, drug-refractory paroxysmal atrial fibrillation (n=142). Variable-loop circular catheter (VLCC) for pulsed field ablation was evaluated on 12-month freedom from atrial arrhythmia recurrence, repeat ablation, procedural failure, or Class I/III antiarrhythmic drug use. First-time pulmonary vein isolation using the variable-loop circular catheter achieved a 12-month primary effectiveness success of 74.8% and a primary adverse event rate of 1.6%.