= 0.999), with the mean slope of 0.0115 (95%CI, 0.0108-0.0122) and intercept of 0.0027 (95%CI, 0.0003-0.0051). The limit of detection (LOD) of the method was 3.17 mg/dL (0.280 mmol/L). Analytical specificity was achieved by chromatographically separating creatinine from potentially interfering creatine within a one-minute run and monitoring the Quantitation Ion/Confirmation Ion (QI/CI) ratios in samples. A simple, accurate, high-throughput method was successfully developed for measuring creatinine in human urine samples.
No takes yet. Share an insight, caveat, or question.
Zahoor et al. (2018) studied this question.