We present a sample preparation method for cryo-electron microscopy (cryo-EM) that requires only 3-20nL of sample to prepare a cryo-EM grid, depending on the protocol used. The sample is applied and spread on the grid by a microcapillary. The procedure does not involve any blotting steps, and real-time monitoring allows the water film thickness to be assessed and decreased to an optimum value prior to vitrification. We demonstrate that the method is suitable for high-resolution cryo-EM and will enable alternative electron microscopy approaches, such as single-cell visual proteomics.
No takes yet. Share an insight, caveat, or question.
Arnold et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: