In this work, we tested the influence of various sampling techniques and modes of operation on the response of sensor systems under real conditions. The first system was based on diffusive sampling. The second one used a dynamic method of sampling. In the third system, a stop flow mode of operation was applied. A considerable error of target gas concentration assessment was obtained in the first two cases. The error resulted from sensor signal fluctuations encountered during their exposure either to the air at the measurement point (diffusive sampling) or to the air drawn from the measurement point (dynamic sampling). The fluctuations could be attributed to the temporal variation of physical and chemical parameters of air. The main merit of the third method consists in utilizing sensor exposure under the conditions of stopped flow for the purpose of pollutant concentration assessment. While using this method, the accuracy of target gas quantification under real conditions was comparable to the one achieved when measuring standard gases. We think that the proposed method is a valuable contribution, which addresses the increasing demand for sensor systems that perform quantitative pollution assessment in workplace air, under real conditions.