Why the study?
Three 64-channel cardiac coils with different detector array configurations were designed to evaluate acceleration capabilities in simultaneous multislice imaging for 3T cardiac MRI.
A 64-channel cardiac coil with non-uniform loop size and density improves highly accelerated simultaneous multislice cardiac MRI compared to symmetrically distributed loop designs.
Guides preclinical cardiac coil optimization; leaves open human translation of non-uniform array benefits.
Purpose Three 64‐channel cardiac coils with different detector array configurations were designed and constructed to evaluate acceleration capabilities in simultaneous multislice (SMS) imaging for 3T cardiac MRI. Methods Three 64‐channel coil array configurations obtained from a simulation‐guided design approach were constructed and systematically evaluated regarding their encoding capabilities for accelerated SMS cardiac acquisitions at 3T. Array configuration A Uni‐sized consists of uniformly distributed equally sized loops in an overlapped arrangement, B Gapped uses a gapped array design with symmetrically distributed equally sized loops, and C Dense has non‐uniform loop density and size, where smaller elements were centered over the heart and larger elements were placed surrounding the target region. To isolate the anatomic variation from differences in the coil configurations, all three array coils were built with identical semi‐adjustable housing segments. The arrays’ performance was compared using bench‐level measurements and imaging performance tests, including signal‐to‐noise ratio (SNR) maps, array element noise correlation, and SMS acceleration capabilities. Additionally, all cardiac array coils were evaluated on a healthy volunteer. Results The array configuration C Dense with the non‐uniformly distributed loop density showed the best overall cardiac imaging performance in both SNR and SMS encoding power, when compared to the other constructed arrays. The diffusion weighted cardiac acquisitions on a healthy volunteer support the favorable accelerated SNR performance of this array configuration. Conclusion Our results indicate that optimized highly parallel cardiac arrays, such as the 64‐channel coil with a non‐uniform loop size and density improve highly accelerated SMS cardiac MRI in comparison to symmetrically distributed loop array designs.
No takes yet. Share an insight, caveat, or question.
Etzel et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: