We have recently examined the potential of bare nanocapillaries for free solution DNA separations and demonstrated efficiencies exceeding 10(6) theoretical plates/m. In the present work, we demonstrate the use of bare and hydroxypropylcellulose (HPC) coated open tubular nanocapillaries for protein separations. Using 1.5 microm inner diameter (i.d.) capillary columns, hydrodynamically injecting femto- to picoliter volumes of fluorescent or fluorescent dye labeled protein samples, utilizing a pneumatically pressurized chamber containing 1.0 mM sodium tetraborate solution eluent (typically 200 psi) as the pump, and performing on-column detection using a simple laser-induced fluorescence detector, we demonstrate efficiencies of close to a million theoretical plates/m while generating single digit microliter volumes of waste for a complete chromatographic run. We achieve baseline resolution for a protein mixture consisting of transferrin, alpha-lactalbumin, insulin, and alpha-2-macroglobulin.