May 18, 2023JACC. Clinical electrophysiology53 citationsOpen Access

Enhanced Thromboresistance and Endothelialization of a Novel Fluoropolymer-Coated Left Atrial Appendage Closure Device

View Full Paper
WSWalid I. SalibaKKKenji KawaiYSYu Sato

Structured PICO

Does a fluoropolymer-coated WATCHMAN FLX device reduce thrombogenicity and improve endothelialization compared to an uncoated device in preclinical models?

P
Population
Canine and porcine models, and flow loop experiments
I
Intervention
Fluoropolymer-coated WATCHMAN FLX (FP-WM) device
C
Comparator
Conventional uncoated WATCHMAN FLX (WM) device
O
Outcome
Thrombogenicity (device-related thrombus, platelet adhesion, albumin adsorption) and endothelial coverage (EC)surrogate

A novel fluoropolymer-coated LAAC device demonstrated reduced thrombogenicity and improved endothelialization in preclinical models compared to the conventional uncoated device.

Abstract

BACKGROUND: Device-related thrombus (DRT) after left atrial appendage closure (LAAC) procedures is a rare but potentially serious event. Thrombogenicity and delayed endothelialization play a role in the development of DRT. Fluorinated polymers are known to have thromboresistant properties that may favorably modulate the healing response to an LAAC device. OBJECTIVES: The goal of this study was to compare the thrombogenicity and endothelial coverage (EC) after LAAC between the conventional uncoated WATCHMAN FLX (WM) and a novel fluoropolymer-coated WATCHMAN FLX (FP-WM). METHODS: Canines were randomized for implantation with WM or FP-WM devices and given no postimplant antithrombotic/antiplatelet agents. The presence of DRT was monitored by using transesophageal echocardiography and verified histologically. The biochemical mechanisms associated with coating were assessed by using flow loop experiments to quantify albumin adsorption, platelet adhesion, and porcine implants to quantify EC and the expression of markers of endothelial maturation (ie, vascular endothelial-cadherin/p120-catenin). RESULTS: ; P = 0.03) and significantly less platelet adhesion (44.7% IQR: 27.2%-60.2% vs 60.9% IQR: 39.9%-70.1%; P < 0.01) on FP-WM. Porcine implants showed significantly greater EC by scanning electron microscopy (87.7% IQR: 83.4%-92.3% vs 68.2% IQR: 47.6%-72.8%; P = 0.03), and higher vascular endothelial-cadherin/p120-catenin expression after 3 months on FP-WM compared with WM. CONCLUSIONS: The FP-WM device showed significantly less thrombus and reduced inflammation in a challenging canine model. Mechanistic studies indicated that the fluoropolymer-coated device binds more albumin, leading to reduced platelet binding, less inflammation, and greater EC.