Why the study?
Does the oscillometric SphygmoCor XCEL device provide equivalent cf-PWV and cSP measurements compared to applanation tonometry in children and adolescents?
Does the oscillometric SphygmoCor XCEL device provide equivalent cf-PWV and cSP measurements compared to applanation tonometry in children and adolescents?
The oscillometric SphygmoCor XCEL device provides equivalent cf-PWV and cSP measurements to applanation tonometry, making it a valid and potentially more practical tool for pediatric populations.
May support oscillometric cf-PWV/cSP use in youth; leaves open prospective validation before practice change.
BACKGROUND: Vascular phenotype by assessing carotid-femoral pulse wave velocity (cf-PWV) and central SBP (cSP) in the young could be used as an intermediate cardiovascular outcome measure. Tonometry is considered the gold-standard technique, but its use is challenging in clinical practice, especially when used in children. The purpose of this study was to validate cf-PWV and cSP assessment with novel oscillometric device (SphygmoCor XCEL) in children and adolescents. METHODS: cf-PWV and cSP were measured in 72 children and adolescents aged 6-20 years. Measurements were performed by applanation tonometry and by the SphygmoCor XCEL device at the same visit under standardized conditions. Regression analysis and Bland-Altman plots were used for comparison of the tonometer-based with oscillometric-based method. RESULTS: Mean cf-PWV measured by applanation tonometry was 4.85 ± 0.81 m/s and measured by SpygmoCor XCEL was 4.75 ± 0.81 m/s. The mean difference between the two devices was 0.09 ± 0.47 m/s (P = NS). cSP measured by SpygmoCor XCEL was strongly correlated with cSP measured by applanation tonometry (R = 0.87, P < 0.001). Mean cSP measured by applanation tonometry was 103.23 ± 9.43 mmHg and measured by SpygmoCor XCEL was 103.54 ± 8.87 mmHg. The mean cSP difference between the two devices was -0.30 ± 3.34 mmHg (P = NS), and fulfilled the AAMI criterion 1. The estimated intersubject variability was 2.17 mmHg. CONCLUSION: The new oscillometric SphygmoCor XCEL device provides equivalent results for cf-PWV and cSP values to those obtained by tonometry in children and adolescents. Thus, the SphygmoCor XCEL device could be appropriate for assessing cf-PWV and cSP in the pediatric population.
No takes yet. Share an insight, caveat, or question.
Stabouli et al. (2018) studied this question.