September 1, 1997Journal of Computer Assisted Tomography221 citations

Comparison Between Manual and Semiautomated Analysis of Left Ventricular Volume Parameters from Short-Axis MR Images

View Full Paper
RGRob J. van der GeestVBV.G.M. BullerEJEric Jansen

Structured PICO

Does a semiautomated contour detection algorithm improve analysis time while maintaining accuracy compared to manual analysis for left ventricular volume parameters in MRI?

P
Population
20 individuals (10 normal volunteers and 10 infarct patients) undergoing short-axis multislice GRE MRI studies
I
Intervention
Semiautomated contour detection algorithm for quantitative analysis of cardiovascular MRI
C
Comparator
Manual image analysis (manually traced contours)
O
Outcome
Left ventricular function parameters (end-diastolic volume, end-systolic volume, ejection fraction, left ventricular mass) and total analysis timesurrogate

Semiautomated contour detection for cardiovascular MRI significantly reduces analysis time while maintaining high accuracy and precision for global left ventricular function parameters.

Abstract

PURPOSE: The goal of this study was to evaluate a newly developed semiautomated contour detection algorithm for the quantitative analysis of cardiovascular MRI. METHOD: Left ventricular function parameters derived from automatically detected endocardial and epicardial contours were compared with results derived from manually traced contours in short-axis multislice GRE MRI studies of 10 normal volunteers and 10 infarct patients. RESULTS: Compared with manual image analysis, the semiautomated method resulted in the following systematic and random differences (auto-manual; mean +/- SD): end-diastolic volume: -5.5 +/- 9.7 ml; end-systolic volume: -3.6 +/- 6.5 ml; ejection fraction: 1.7 +/- 4.1%; left ventricular mass: 7.3 +/- 20.6 g. Total analysis time for a complete study was reduced from 3-4 h for the manual analysis to < 20 min using semiautomated contour detection. CONCLUSION: Global left ventricular function parameters can be obtained with a high degree of accuracy and precision using the present semiautomated contour detection algorithm.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Geest et al. (1997) studied this question.

synapsesocial.com/papers/6a1d3a70750575be8d2f4e15https://doi.org/10.1097/00004728-199709000-00019