January 1, 2000Magnetic Resonance in Medicine578 citationsOpen Access

Prospective acquisition correction for head motion with image-based tracking for real-time fMRI

View Full Paper
STStefan ThesenOHO. HeidEMEdgar A. Mueller

Key Points

Key points are not available for this paper at this time.

Abstract

In functional magnetic resonance imaging (fMRI) head motion can corrupt the signal changes induced by brain activation. This paper describes a novel technique called Prospective Acquisition CorrEction (PACE) for reducing motion-induced effects on magnetization history. Full three-dimensional rigid body estimation of head movement is obtained by image-based motion detection to a high level of accuracy. Adjustment of slice position and orientation, as well as regridding of residual volume to volume motion, is performed in real-time during data acquisition. Phantom experiments demonstrate a high level of consistency (translation < 40 microm; rotation < 0.05 degrees ) for detected motion parameters. In vivo experiments were carried out and they showed a significant decrease of variance between successively acquired datasets compared to retrospective correction algorithms.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Thesen et al. (2000) studied this question.

synapsesocial.com/papers/6a1d44277f448865515e15b3https://doi.org/10.1002/1522-2594(200009)44:3<457::aid-mrm17>3.0.co;2-r