Why the study?
Does DDDR pacing with mode switch improve quality of life compared to DDIR pacing in patients with complete heart block and paroxysmal atrial tachyarrhythmias?
Does DDDR pacing with mode switch improve quality of life compared to DDIR pacing in patients with complete heart block and paroxysmal atrial tachyarrhythmias?
In patients with complete heart block and paroxysmal atrial tachyarrhythmias, DDDR pacing with mode switch preserves AV synchrony better and is preferred by most patients compared to DDIR pacing, despite similar overall quality of life scores.
Answers are grounded in this paper and cite their sources.
DDDR may be favored by patient preference despite equivalent QOL; extends pacing mode comparisons in heart block with paroxysmal atrial tachyarrhythmias.
A prospective double-blind randomized crossover study was done in 15 patients with complete heart block and intermittent ATs. The pacemaker was randomly programmed to dual chamber inhibited rate responsive pacing (DDIR) and to DDDR with mode switch, for 1 month each. An event recorder was given to the patients and after each period, a QOL questionnaire was obtained. Based on telemetric data, all but two patients had AT during follow-up. The duration and frequency of these episodes were not related to mode settings. AV synchrony was better preserved in DDDR (P < 0.05). Most symptom-related event recordings during DDIR showed loss of AV synchrony; DDDR with mode switch caused symptoms due to tracking of ST. Overall the QOL score was not different between the modes. Fewer somatic complaints were noted during DDDR pacing than during baseline. DDIR stimulation showed no difference. Twelve patients preferred the period of DDDR pacing; one experienced severe symptoms during DDIR. In conclusion, patients with paroxysmal AT, DDDR with mode switch, and DDIR had no influence on the occurrence, nor on the duration of AT episodes. AV synchrony was better preserved in DDDR, which was also associated with fewer somatic complaints compared to the baseline. In DDDR, symptoms were observed when ST was tracked. QOL was comparable, although more patients preferred DDDR.
No takes yet. Share an insight, caveat, or question.
Provenier et al. (1999) studied this question.