BACKGROUND: The Afinion AS100 analyser is a small bench-top, multi-assay, point-of-care (POC) analyser that is able to measure glycated haemoglobin (HbA1c) and lipid levels. OBJECTIVE: To assess performance of the Afinion analyser compared with a reference laboratory test for the measurement of HbA1c and lipid levels. METHOD: The study involved men and women enrolled in a cross-sectional study, Sexual health, HIV infection and comorbidity with non-communicable diseases among Older Persons (SHIOP), which was conducted from February to May 2016. Whole blood was drawn aseptically by a trained study nurse into a serum separator gel tube and an ethylenediaminetetra-acetic acid (EDTA) tube. The EDTA whole blood was used to measure HbA1c levels, and serum to measure total cholesterol (TC), high-density lipoprotein cholesterol (HDL-C), low-density lipoprotein cholesterol (LDL-C) and triglyceride levels. Lin's correlation coefficient was used to assess the agreement between the Afinion and ABX Pentra 400 analysers for each marker. RESULTS: A total of 435 older individuals were included in the study. The proportion of HbA1c results that were correctly classified by the Afinion analyser was 92.2%. Bland-Altman analysis and linear regression analysis showed a very good agreement (correlation concordance 0.89) between the two analysers for the measurement of HbA1c. The two-way scatter plot for TC showed a substantial correlation (0.80). However, a total of 69 cholesterol results that were within the normal range on the Pentra were misclassified as abnormal on the Afinion. The readings obtained for HDL-C levels with the Afinion were shown to be slightly overestimated when compared with the Pentra. However, correlation for HDL-C on the two analysers was 0.93, indicating an almost perfect agreement. Seventy-four LDL-C results were erroneously classified as abnormal on the Afinion but were within the normal range on the Pentra, resulting in a substantial correlation of 0.75. An excellent agreement was observed between triglyceride measurements (0.99). CONCLUSION: This study supports the use of the Afinion AS100 analyser in POC testing for the measurement of HbA1c, triglycerides and HDL-C in a South African setting.
No takes yet. Share an insight, caveat, or question.
Abbai et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: