Key result
Echocardiographic measures of ejection fraction yielded mean values 5% higher than radionuclide ventriculography (30% vs 25%, p=0.03), with poor agreement that deteriorates with ventricular pacing.
Why the study?
How do different echocardiographic methods compare with radionuclide ventriculography for measuring left ventricular function in patients with permanent pacemakers?
Observational (n=307)
How do different echocardiographic methods compare with radionuclide ventriculography for measuring left ventricular function in patients with permanent pacemakers?
Echocardiographic and radionuclide ventriculography measurements of left ventricular function are poorly interchangeable in patients with pacemakers, and accuracy depends on the underlying rhythm.
Echocardiographic and radionuclide LVEF values should not be interchanged in pacemaker patients; leaves open optimal modality selection for serial monitoring.
AIMS: Different methods exist for measuring left ventricular function echocardiographically; each may be error prone due to the abnormal pattern of ventricular activation during pacing. METHODS AND RESULTS: Echocardiography was undertaken on 307 patients with permanent pacemakers; a subset of 57 underwent radionuclide ventriculography. Intrinsic and paced beats were analysed for left ventricular function by: Simpson's bi-plane, Teicholz M-mode, wall-motion scoring and 'eyeball' assessment. Agreement between techniques and with radionuclide ventriculography were compared according to intrinsic or paced beats. Echocardiographic measures of ejection fraction give mean values 5% higher than radionuclide ventriculography (Simpson's 30+/-9%, vs. Teicholz 30+/-13% vs. radionuclide ventriculography 25+/-9%, p=0.03). Agreement between Simpson's, Teicholz and radionuclide ventriculography by Bland-Altman analysis showed poor agreement (Simpson's vs. Teicholz range (4xSD)=57%, Simpson's vs. radionuclide ventriculography=36%, Teicholz vs. radionuclide ventriculography=46%, p=0.02), the level of agreement deteriorates with ventricular pacing (Simpson's vs. Teicholz range=61%, Simpson's vs. radionuclide ventriculography=34%, Teicholz vs. radionuclide ventriculography=47%, p=0.02). The correlation between wall motion analysis and radionuclide ventriculography is moderately poor (all subjects r=0.58, ventricular pacing r=0.52, not pacing r=0.66). CONCLUSION: Echocardiography and radionuclide ventriculography are the only non-invasive techniques to assess left ventricular function in the paced population. Results are poorly interchangeable and the accuracy of any comparison dependent on the underlying rhythm.
No takes yet. Share an insight, caveat, or question.
Thackray et al. (2005) conducted an observational in Patients with permanent pacemakers (n=307). Echocardiography vs. Radionuclide ventriculography was evaluated on Agreement between echocardiographic techniques and radionuclide ventriculography for left ventricular function (p=0.03). Echocardiographic measures of ejection fraction yielded mean values 5% higher than radionuclide ventriculography (30% vs 25%, p=0.03), with poor agreement that deteriorates with ventricular pacing.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: