Key result
A novel portable glucose monitoring system demonstrated a sensitivity of 1.18 Hz/mM·cm2 and a correlation coefficient of 0.9939 for glucose concentrations from 1 mM to 25 mM.
A novel, cost-effective portable glucose monitoring system using a biofuel cell demonstrated high sensitivity and correlation for glucose concentrations in physiological ranges.
Answers are grounded in this paper and cite their sources.
May advance portable glucose sensor development; leaves open human validation before clinical use.
A cost-effective portable glucose monitoring system with remote data access based on a novel e-oscilloscope was developed using a glucose biofuel cell and a capacitor circuit interfaced to an ESP8266 microcontroller programmed to convert the charge/discharge rates of the capacitor functioning as a transducer. The capacitor charge/discharge rates were converted into glucose concentration readings that is monitored remotely. The glucose monitoring system comprise a glucose biofuel cell, a charge pump circuit, a capacitor and an ESP microcontroller. The anode was fabricated by modifying a gold microwire with nanoporous colloidal platinum (Au-co-Pt) and the cathode was constructed using a mesh dense network of multiwalled carbon nanotubes modified with bilirubin oxidase, respectively. The glucose monitoring system showed sensitivity of 1.18 Hz/mM·cm 2 with a correlation coefficient of 0.9939 with increasing glucose concentration from 1 mM to 25 mM. In addition, the glucose monitoring system exhibited optimal operation at a pH of 7.4 and 37 °C, which is ideal for physiological glucose monitoring.
No takes yet. Share an insight, caveat, or question.
Baingane et al. (2020) studied Glucose monitoring. Glucose monitoring system with remote data access was evaluated on Sensitivity and correlation coefficient (Correlation coefficient 0.9939). A novel portable glucose monitoring system demonstrated a sensitivity of 1.18 Hz/mM·cm2 and a correlation coefficient of 0.9939 for glucose concentrations from 1 mM to 25 mM.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: