Key result
Auto-EF yielded higher inter-operator correlation (r=0.94 vs r=0.70, P<0.01) and narrower limits of agreement than the Simpson's method for assessing LVEF in critically ill patients.
Why the study?
Does Auto-EF improve the reproducibility of LVEF assessment between inexperienced and experienced operators compared to the biplane Simpson's method in critically ill patients?
Observational (n=29)
Does Auto-EF improve the reproducibility of LVEF assessment between inexperienced and experienced operators compared to the biplane Simpson's method in critically ill patients?
Auto-EF provides more reproducible LVEF measurements than the traditional Simpson's method for inexperienced operators evaluating critically ill patients.
Answers are grounded in this paper and cite their sources.
Auto-EF may improve LVEF reproducibility between operators in the ICU; hypothesis-generating and requires prospective validation before practice change.
BACKGROUND: The biplane Simpson's method is considered the gold standard to assess and monitor left ventricular (LV) ejection fraction (EF) in critically ill patients. Recently, a new semi-automatic technique based on speckle tracking echocardiography called "Auto-EF" has been introduced. We compared LVEF values obtained with biplane Simpson's method and Auto-EF by two groups of operators: trainee echocardiography intensivists and experienced echocardiographers. METHODS: A standard transthoracic echocardiography was performed on 37 patients. According to image quality 29 patients were selected. Each inexperienced and experienced operator executed an off-line analysis using both Simpson's method and Auto-EF. LVEF obtained by the two groups of operators were then compared. RESULTS: EF values assessed with Simpson's method showed a moderate correlation (r=0.70, P<0.01) between inexperienced and experienced operators. The Bland-Altman analysis showed a mean bias of 0.3% with limits of agreement (LoA) from -24.4 to +25.1%. Values obtained with Auto-EF showed a good correlation (r=0.94, P<0.01) with a mean bias of 0.2% and LoA from -10.1 to +10.4%. CONCLUSIONS: Due to its semiautomatic nature, for inexpert operators Auto-EF seems more reproducible than the traditional Simpson's method for monitoring left ventricular function in critically ill patients.
No takes yet. Share an insight, caveat, or question.
Franchi et al. (2018) conducted an observational in Critically ill patients (n=29). Auto-EF (semi-automatic speckle tracking echocardiography) vs. Biplane Simpson's method was evaluated on Inter-operator correlation of LVEF values between inexperienced and experienced operators (r=0.94 vs r=0.70, p=<0.01). Auto-EF yielded higher inter-operator correlation (r=0.94 vs r=0.70, P<0.01) and narrower limits of agreement than the Simpson's method for assessing LVEF in critically ill patients.