Cholesterol oxidase has been immobililzed on collagen films associated to an electrochemical sensor to form a “cholesterol electrode”. The electrode poised at a potential of + 650 my vs Ag/AgCl detects the hydrogen peroxide produced in the enzymatic reaction. This device presents a very high sensitivity and a wide range of linearity (10−7 M - 0.8.10−4 M). The use of a non enzymatic electrode associated with the enzymatic one allowed the detection and correction of electrochemical interferences when applied to human sera for free cholesterol determination.
No takes yet. Share an insight, caveat, or question.
Bertrand et al. (1979) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: