Key result
Supreme DES matches DP-EES for 1-year target lesion failure across ACS and CCS.
Why the study?
The relative safety and effectiveness of the Supreme biodegradable polymer DES in patients with acute coronary syndromes compared with chronic coronary syndromes was unknown.
Does the Supreme DES improve target lesion failure at 12 months compared to DP-EES in patients with acute and chronic coronary syndromes?
RCT (n=1,628)
Single-blind
2:1
Yes
Does the Supreme DES improve target lesion failure at 12 months compared to DP-EES in patients with acute and chronic coronary syndromes?
Absolute Event Rate: 6.6% vs 6%
p-value: p=0.89
The Supreme biodegradable polymer sirolimus-eluting stent demonstrated similar 12-month safety and efficacy compared to durable polymer everolimus-eluting stents in both acute and chronic coronary syndrome patients.
Supreme DES offers a comparable option to DP-EES in ACS and CCS; reinforces randomized evidence for biodegradable-polymer stents.
Abstract Background The PIONEER III trial demonstrated noninferiority of 12-month target lesion failure (TLF) with the Supreme DES (Sinomed), a thin-strut cobalt-chromium, biodegradable polymer, sirolimus-eluting stent, compared with a durable polymer, everolimus-eluting (XIENCE/PROMUS) stent (DP-EES). The relative safety and effectiveness of the Supreme DES in patients with acute coronary syndromes (ACS) and those with chronic coronary syndromes (CCS) is not known. Methods PIONEER III was a prospective, multicenter, international, 2:1 randomized trial stratified by clinical presentation. The primary end point was TLF at 12 months (a composite of cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization). Results A total of 1628 patients were enrolled, including 41% of patients with ACS (unstable angina and non–ST-elevation myocardial infarction) randomized to Supreme DES (n = 441) versus DP-EES (n = 232) and 59% of patients with CCS randomized to Supreme DES (n = 645) versus DP-EES (n = 310). Patients with ACS were younger, fewer presented with less diabetes, hypertension, and previous revascularization, but more were current smokers. The primary end point of TLF (6.4% vs 4.4%; P = .1), major adverse cardiac events (8.5% vs 6.5%; P = .16), and stent thrombosis (0.4% vs 0.9%; P = .25) at 12 months were similar in the ACS and CCS groups. There was no difference in TLF at 12 months between Supreme DES and DP-EES among patients with ACS (6.6% vs 6.0%; P = .89) and those with CCS (4.5% vs 4.3%; P = .83); interaction P = .51 for TLF by clinical presentation. Conclusions Compared with the DP-EES, the Supreme DES seemed safe and effective with a similar TLF at 12 months in both patients with ACS and those with CCS.
No takes yet. Share an insight, caveat, or question.
Hussain et al. (2023) conducted an RCT in Acute and Chronic Coronary Syndromes (n=1,628). Supreme DES vs. Durable polymer everolimus-eluting stent (DP-EES) was evaluated on Target lesion failure (TLF) at 12 months in patients with ACS (p=0.89). Supreme DES demonstrated similar rates of target lesion failure at 12 months compared to DP-EES in both patients with acute coronary syndromes (6.6% vs 6.0%, p=0.89) and chronic coronary syndromes (4.5% vs 4.3%, p=0.83).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: