Key result
The Palmaz-Schatz-shaped stent exhibited the least dogboning and recoiling with favorable hemodynamics, while the modified C-Rcrown stent improved upon the Cypher-shaped stent's performance.
Why the study?
Different coronary stent structures have various impacts on arterial and plaque stress distribution as well as the local hemodynamic environment.
Does stent structure design affect mechanical stress and local hemodynamics in a curved stenotic coronary artery model?
Does stent structure design affect mechanical stress and local hemodynamics in a curved stenotic coronary artery model?
Computational modeling demonstrates that stent structure significantly impacts both mechanical stress distribution and local hemodynamics, with specific designs like C-Rcrown and C-Astrut showing improved performance over standard Cypher-shaped stents.
Stent geometry optimization remains premature for clinical use; leaves open need for human trials to validate hemodynamic advantages.
Coronary artery stenting is commonly used for the treatment of coronary stenosis, and different stent structures indeed have various impacts on the stress distribution within the plaque and artery as well as the local hemodynamic environment. This study aims to evaluate the performance of different stent structures by characterizing the mechanical parameters after coronary stenting. Six stent structures including three commercially-shaped stents (Palmaz-Schatz-shaped, Xience Prime-shaped, and Cypher-shaped) and three author-developed stents (C-Rlink, C-Rcrown, and C-Astrut) implanted into a curved stenotic coronary artery were investigated. Structural analyses of the balloon-stent-plaque-artery system were first performed, and then followed by hemodynamic analyses. The results showed that among the three commercially-shaped stents, the Palmaz-Schatz-shaped had the least stent dogboning and recoiling, corresponding to the greatest maximum plastic strain and the largest diameter change, nevertheless, it induced the highest maximum von Mises stress on plaque, arterial intima and media. From the viewpoint of hemodynamics, the Palmaz-Schatz-shaped displayed smaller areas of adverse low wall shear stress (<0.5 Pa), low time-averaged wall shear stress (<0.5 Pa), and high oscillating shear index (>0.1). Compared to the Cypher-shaped, the C-Rcrown and C-Astrut had smaller recoiling, greater maximum plastic stain and larger diameter change, which indicated the improved mechanical performance of the Cypher-shaped stent. Moreover, both C-Rcrown and C-Astrut exhibited smaller areas of adverse low wall shear stress, and low time-averaged wall shear stress, but only the C-Rcrown displayed a smaller area of adverse high oscillating shear index. The present study evaluated and compared the performance of six different stents deployed inside a curved artery, and could be potentially utilized as a guide for the selection of suitable commercially-shaped stent for clinical application, and to provide an approach to improve the performance of the commercial stents.
No takes yet. Share an insight, caveat, or question.
Wei et al. (2019) studied Coronary artery stenosis. Stent structures (Palmaz-Schatz-shaped, Xience Prime-shaped, Cypher-shaped, C-Rlink, C-Rcrown, C-Astrut) vs. Comparison among different stent designs was evaluated on Structural (dogboning, recoiling, stress) and hemodynamic (WSS, TAWSS, OSI) parameters. The Palmaz-Schatz-shaped stent exhibited the least dogboning and recoiling with favorable hemodynamics, while the modified C-Rcrown stent improved upon the Cypher-shaped stent's performance.