Key result
Head and lower lumbar 3 T MRI scans in patients with Biotronik ProMRI ICD systems were safe, with a 100% freedom from MRI-related serious adverse device effects (95% CI 95.98-100%).
Why the study?
There were no published studies evaluating the safety of 3 T MRI in patients with MRI-conditional ICDs.
Does 3 Tesla MRI cause serious adverse device effects or alter pacing/sensing parameters in patients with MRI-conditional implantable cardioverter-defibrillators?
Observational (n=129)
Yes
Does 3 Tesla MRI cause serious adverse device effects or alter pacing/sensing parameters in patients with MRI-conditional implantable cardioverter-defibrillators?
Head and lower lumbar 3 T MRI scans are safe in patients with specific MRI-conditional ICD systems, with no adverse device effects or significant changes in pacing/sensing parameters.
Answers are grounded in this paper and cite their sources.
Supports 3T MRI in Biotronik ProMRI ICDs; extends observational safety data but leaves open need for randomized confirmation.
AIMS: There have been no published studies on the safety of magnetic resonance imaging (MRI) at 3 Tesla (3 T) in patients with MRI-conditional implantable cardioverter-defibrillators (ICDs). The aim of this study was to assess clinical safety of the Biotronik ProMRI ICD system during non-diagnostic head and lower lumbar scans under 3 T MRI conditions. METHODS AND RESULTS: The study enrolled 129 patients at 12 sites in Australia, Singapore, and Europe. Predefined head and lower lumbar MR scans (total duration ≈30 min) were performed in 112 patients. Three primary endpoints were evaluated from the pre-MRI to the 1-month post-MRI visit: (i) freedom from serious adverse device effects (SADEs) related to MRI (hypothesized to be >90%); (ii) pacing threshold invariance for all leads (geometric mean of the patient-wise ratios for 1 month vs. pre-MRI was hypothesized to be <1.07); and (iii) sensing amplitude invariance (geometric mean of the ratios was hypothesized to be >0.993). No MRI-related SADE occurred (SADE-free rate 100%, 95% confidence interval 95.98-100%). Pacing threshold and sensing amplitudes fulfilled the invariance hypotheses with high statistical significance (P < 0.0013). No threshold increase >0.5 V or sensing amplitude decrease by >50% was observed (secondary endpoints). Lead impedances, battery capacity, and detection and treatment of arrhythmias by ICDs were not affected by MRI scans. CONCLUSION: The head and lower lumbar scans under specific 3 T MRI conditions were safe in the investigated MR-conditional ICD systems. There was no evidence of harm to the patients or any negative influence of the MRI scan on the implanted systems.
No takes yet. Share an insight, caveat, or question.
Zbinden et al. (2019) conducted an observational in MRI-conditional implantable cardioverter-defibrillators (n=129). 3 T MRI head and lower lumbar scans was evaluated on Freedom from serious adverse device effects (SADEs) related to MRI (95% CI 95.98-100). Head and lower lumbar 3 T MRI scans in patients with Biotronik ProMRI ICD systems were safe, with a 100% freedom from MRI-related serious adverse device effects (95% CI 95.98-100%).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: