Answers are grounded in this paper and cite their sources.
Purpose To investigate the effect of stretching sampling window on quantitative 3D ultrashort TE (UTE) imaging of cortical bone at 3 T. Methods Ten bovine cortical bone and 17 human tibial midshaft samples were imaged with a 3T clinical MRI scanner using an 8‐channel knee coil. Quantitative 3D UTE imaging biomarkers, including T 1 , , magnetization transfer ratio and magnetization transfer modeling, were performed using radial or spiral Cones sampling trajectories with various durations. Errors in UTE‐MRI biomarkers as a function of sampling time were evaluated using radial sampling as a reference standard. Results For both bovine and human cortical bone samples, no significant differences were observed for all UTE biomarkers (single‐component , bicomponent and relative fractions, T 1 , magnetization transfer ratio, and magnetization transfer modeling of macromolecular fraction) for spiral sampling windows of 992 µs to 1600 µs compared with a radial sampling window of 688 µs. Conclusion The total scan time can be reduced by 76% with quantification errors less than 5%. Quantitative UTE‐MRI techniques can be greatly accelerated using longer sampling windows without significant quantification errors.
No takes yet. Share an insight, caveat, or question.
Wan et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: