Why the study?
Due to limited safety data, many non-pacemaker/ICD metallic devices and abandoned leads continue to be considered contraindications for MRI.
Is MRI safe and feasible in patients with non-pacemaker/defibrillator metallic devices and abandoned leads when using a restricted SAR and scan time protocol?
Cohort (n=94)
No
Is MRI safe and feasible in patients with non-pacemaker/defibrillator metallic devices and abandoned leads when using a restricted SAR and scan time protocol?
MRI at 1.5 Tesla using a controlled protocol (SAR < 4.0 W/kg, scan time < 60 minutes) appears feasible and safe in patients with non-pacemaker/defibrillator metallic devices and abandoned leads.
OBJECTIVE: To evaluate feasibility of MRI in patients with non-pacemaker (PM)/ Implantable cardioverter defibrillator (ICD) metallic devices and abandoned leads. BACKGROUND: Relative safety of MRI performed using specified protocol has been established in MR non-conditional PM/ICDs. With limited safety data, many non-PM/ICD metallic devices and abandoned leads continue to be a contraindication for MRI. METHODS: We retrospectively analyzed consecutive patients with extra-cardiac devices, non-programmable cardiac devices, and abandoned leads, who underwent MRI (GE 1.5 Tesla, WI) at a single tertiary care center over a span of 13 years. Scan protocol was designed to maintain specific absorption rate (SAR) < 4.0 W/kg and scan time < 60 minutes. RESULTS: The cohort comprised 127 MRI exams representing 94 patients, with 13 patients having two or more scans. The devices consisted of: 23 vagal nerve stimulators (VNS), 22 implantable loop recorders, 16 spinal stimulators, 5 peripheral nerve stimulators, 3 bladder stimulators, 2 deep brain stimulators, 1 gastric stimulator, 1 bone stimulator, 1 WATCHMAN device, 22 abandoned PM/lCD leads and 1 VNS lead. There was no immediate (peri-MRI exam) morbidity or mortality. Patients did not report any discomfort, palpitations, heating, or sensation of device migration during the exam. Local follow-up data was available in 65% (100% for thoracic imaging) with a mean of 190±475 days (median 13 days). No device malfunction reported during follow-up. CONCLUSIONS: With appropriate precautions, MRI is feasible in patients with extracardiac devices, nonprogrammable cardiac devices, and abandoned leads.
No takes yet. Share an insight, caveat, or question.
Gopalakrishnan et al. (2021) conducted a cohort in Non-pacemaker/defibrillator metallic devices and abandoned leads (n=94). MRI (1.5 Tesla) was evaluated on Immediate morbidity, mortality, and device malfunction.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: