Radiology-based lung-cancer detection is a high-contrast imaging task, consisting of the detection of a small mass of tissue within much lower density lung parenchyma. This imaging task requires removal of confounding image details, fast image acquisition (< 0.1 s for pericardial region), low dose (comparable to a chest x-ray), high resolution (< 0.25 mm in-plane) and patient positioning flexibility. We present an investigation of tomosynthesis, implemented using the Scanning-Beam Digital X-ray System (SBDX), to achieve these goals. We designed an image-based computer model of tomosynthesis using a high-resolution (0.15-mm isotropic voxels), low-noise CT volume image of a lung phantom, numerically added spherical lesions and convolution-based tomographic blurring. Lesion visibility was examined as a function of half-tomographic angle for 2.5 and 4.0 mm diameter lesions. Gaussian distributed noise was added to the projected images. For lesions 2.5 mm and 4.0 mm in diameter, half-tomographic angles of at least 6° and 9° respectively were necessary before visualization of the lesions improved. The addition of noise for a dose equivalent to 1/10 that used for a standard chest radiograph did not significantly impair lesion detection. The results are promising, indicating that lung-cancer detection using a modified SBDX system is possible.
No takes yet. Share an insight, caveat, or question.
Fahrig et al. (2003) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: