An advanced Scanning-Beam Digital X-ray (SBDX) system for cardiac angiography has been constructed. The 15-kW source operates at 70 - 120 kVp and has an electron beam that is electromagnetically scanned across a 23-cm X 23-cm transmission target. The target is directly liquid cooled for continuous full-power operation and is located behind a focused source collimator. The collimator is a rectangular grid of 100 X 100 apertures whose axes are aligned with the center of the detector array. X-ray beam divergence through the collimator apertures is matched to the 5.4-cm X 5.4 cm detector, which is 150 cm from the source. The detector is a 48 X 48 element CdZnTe direct-conversion photon-counting detector. A narrow x-ray beam scans the full field of view at up to 30 frames per second. A custom digital processor simultaneously reconstructs sixteen 1,000² pixel tomographic images in real time. The slices are spaced 1.2 cm apart and cover the entire cardiac anatomy. The small detector area and large patient-detector distance result in negligible detected x-ray scatter. Image signal-to-noise ratio is calculated to be equal to conventional fluoroscopic systems at only 12% of the patient exposure and 25% of the staff exposure. Exposure reduction is achieved by elimination of detected scatter, elimination of the anti-scatter grid, increased detector DQE, and increased patient entrance area.