Electron Monochromator Design

View Full Paper
CKC. E. KuyattJSJ. Arol Simpson

Key Points

Key points are not available for this paper at this time.

Abstract

A study has been made of all the known factors which limit the performance of high resolution (0.07 to 0.01 eV FWHM) monochromators. These limiting factors have been incorporated into design equations for the optimum (maximum current output) monochromator. The conclusions are tested by performance measurements on a prototype instrument. The results require the introduction into the design equation of a new limiting factor, an anomalous energy spread in dense electron beams, which is empirically determined.