Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs 90207 device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patients

View Full Paper
Ask AI
Bookmark
Share

Key result

Patients with insulin-dependent diabetes mellitus had a more severe overestimation of systolic blood pressure by the SpaceLabs 90207 device compared to healthy controls (1.7 mmHg difference, P=0.009).

Why the study?

Does the agreement between the SpaceLabs 90207 oscillometric device and the sphygmomanometer differ between insulin-dependent diabetic patients and healthy controls?

Population

106 patients with insulin-dependent diabetes mellitus and 92 healthy control subjects

Comparison

Blood pressure measurement using the SpaceLabs… vs Blood pressure measurement using a…

Design

Cross-sectional

Authors

FIFrans J. van IttersumAmsterdam University Medical CentersRWRenate M. J. WijeringLeiden University Medical CenterJLJ LambertIsala

Discussion

Loading...

Member takes

Implication

May warrant caution interpreting oscillometric BP readings in insulin-dependent diabetes; leaves open need for subpopulation-specific device validation.

Study Design

Type

Observational (n=198)

Structured PICO

Does the agreement between the SpaceLabs 90207 oscillometric device and the sphygmomanometer differ between insulin-dependent diabetic patients and healthy controls?

P
Population
198 participants, comprising 106 patients with insulin-dependent diabetes mellitus and 92 healthy controls, evaluated retrospectively for blood pressure measurement agreement between devices.
E
Exposure
Blood pressure measurement using the SpaceLabs 90207 oscillometric device
C
Comparator
Blood pressure measurement using a sphygmomanometer
O
Outcome
Limits of agreement between the SpaceLabs 90207 device and the sphygmomanometer for systolic, mean arterial, and diastolic blood pressuresurrogate

Main Result

Mean Difference: 1.7

p-value: p=0.009

The limits of agreement between automated oscillometric blood pressure devices and sphygmomanometers differ significantly between diabetic patients and healthy controls, suggesting that device validation protocols should specifically evaluate such subpopulations.

Cite This Study

Ittersum et al. (1998) conducted an observational in Insulin-dependent diabetes mellitus (n=198). Insulin-dependent diabetes mellitus (IDDM) vs. Healthy control subjects was evaluated on Difference in overestimation of systolic blood pressure between the SpaceLabs 90207 device and sphygmomanometer (MD 1.7 mmHg, p=0.009). Patients with insulin-dependent diabetes mellitus had a more severe overestimation of systolic blood pressure by the SpaceLabs 90207 device compared to healthy controls (1.7 mmHg difference, P=0.009).

synapsesocial.com/papers/6a20dcf134bef10fdaeb143ehttps://doi.org/10.1097/00004872-199816080-00007
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Comparative accuracy of six ambulatory devices according to blood pressure levels1993 · 63 citations
  2. 2Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol1991 · 453 citations
  3. 3Sodium, Blood Pressure, and Arterial Distensibility in Insulin-Dependent Diabetes Mellitus1997 · 30 citations
  4. 4The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems1990 · 778 citations