Key result
The proposed automatic 3D atrial wall thickness measurement algorithm achieved less than 10% error compared to manual methods while reducing detection time to 27.15 seconds per patient.
Why the study?
Accurate detection of endocardial and epicardial boundaries remains a prerequisite for automated cardiac wall thickness measurements despite advances in CTA-based segmentation.
Does a novel automated 3D AWT measurement algorithm using PDE improve the efficiency of calculating AWT while maintaining accuracy compared to manual measurement in patients with atrial fibrillation?
Does a novel automated 3D AWT measurement algorithm using PDE improve the efficiency of calculating AWT while maintaining accuracy compared to manual measurement in patients with atrial fibrillation?
The proposed automated 3D atrial wall thickness measurement algorithm significantly reduces processing time to 27 seconds per patient while maintaining high accuracy compared to manual methods.
Despite advancements in high-precision segmentation technology for computed tomographic angiography (CTA)-based cardiac wall segmentation, the accurate detection of the endocardial (Endo) and epicardial (Epi) boundaries remains a prerequisite for automated measurements of the cardiac wall thickness (WT). We proposed a novel algorithm for automated three-dimensional (3D) atrial WT (AWT) measurements, including an automatic Endo-Epi boundary detection. We detected the boundaries that were topologically indistinguishable due to an open geometry at the anatomical boundaries using the combined Convex hull and Poisson solver methods. The Laplace equation for the WT measurement was solved by a partial differential equation combining the two detected boundaries of the myocardial wall. We verified the robustness of our algorithm in mask images of the atrial wall that were separated from the CTA images of 20 patients and a phantom model. The accuracy of the automatically detected Endo-Epi boundaries was acceptable as compared to that manually extracted from the phantom model (Dice coefficient = 0.979). The 3D AWTs calculated by the novel automated method from the CTA images obtained from 20 patients with atrial fibrillation had <10% error as compared to the conventional manual AWT measurement method for comparing the regional WT in all locations except the left antral area. The proposed algorithm’s AWT detection time was 27.15±6.99 s per patient, which was 1/40 that of the conventional method. Consequently, our results showed that the proposed automatic 3D AWT measurement algorithm had the potential to significantly improve the efficiency of calculating the AWT while maintaining the existing level of accuracy.
No takes yet. Share an insight, caveat, or question.
Kwon et al. (2022) studied Atrial fibrillation (n=20). Automated 3D atrial wall thickness measurement algorithm (PDE-based) vs. Conventional manual measurement method was evaluated on Atrial wall thickness detection time (seconds per patient). The proposed automatic 3D atrial wall thickness measurement algorithm achieved less than 10% error compared to manual methods while reducing detection time to 27.15 seconds per patient.