In this paper, we discuss recent advances in the high gain, very large area silicon avalanche photodiode technology. Very large area APDs (square and octagonal design, up to 45 cm/sup 2/ area) have been built using the planar process. These APDs have been successfully operated as scintillation spectrometers. When cooled to -40 /spl deg/C, the energy resolution of 5.9 keV X-rays (/sup 55/Fe source) upon direct interaction with the 45 cm/sup 2/ APD was measured to be 2.4 keV (FWHM). The APD was coupled to a CsI(Tl) scintillator and irradiated with 662 keV gamma rays from /sup 137/Cs source and the energy resolution was measured to be /spl sim/10% (FWHM) at the temperature of -40 /spl deg/C. A rise time of < 2 ns was measured with the 40 cm/sup 2/ APD upon irradiation with 5.5 MeV alpha particles. A new packaging design has been developed to allow cooling of these large APDs to liquid nitrogen temperature. Evaluation of the 45 cm/sup 2/ APD has been conducted at LN/sub 2/ temperatures. Its gain was found to be as high as 10/sup 4/, while the noise was measured to be /spl sim/0.8 electrons (rms). Detection of low intensity optical signal (<8 photons/pulse) with the 45 cm/sup 2/ APD has also been accomplished.