Key result
Anatomy-driven shim region-of-interest selection and cardiac phase-specific shimming improved B0-field homogeneity compared with standard vendor-supplied shimming for human heart MRI at 7T.
Why the study?
Static magnetic B0 field inhomogeneities cause signal loss and image distortions in cardiac MRI at ultrahigh field (>= 7T), requiring effective B0 shimming.
Cardiac phase-specific shimming and anatomy-driven ROI selection improve B0-field homogeneity for human heart MRI at 7T compared to standard vendor-supplied methods.
May enhance 7T cardiac MRI quality; leaves open prospective validation before clinical adoption.
Purpose Inhomogeneities of the static magnetic B 0 field are a major limiting factor in cardiac MRI at ultrahigh field (≥ 7T), as they result in signal loss and image distortions. Different magnetic susceptibilities of the myocardium and surrounding tissue in combination with cardiac motion lead to strong spatio‐temporal B 0 ‐field inhomogeneities, and their homogenization (B 0 shimming) is a prerequisite. Limitations of state‐of‐the‐art shimming are described, regional B 0 variations are measured, and a methodology for spherical harmonics shimming of the B 0 field within the human myocardium is proposed. Methods The spatial B 0 ‐field distribution in the heart was analyzed as well as temporal B 0 ‐field variations in the myocardium over the cardiac cycle. Different shim region‐of‐interest selections were compared, and hardware limitations of spherical harmonics B 0 shimming were evaluated by calibration‐based B 0 ‐field modeling. The role of third‐order spherical harmonics terms was analyzed as well as potential benefits from cardiac phase–specific shimming. Results The strongest B 0 ‐field inhomogeneities were observed in localized spots within the left‐ventricular and right‐ventricular myocardium and varied between systolic and diastolic cardiac phases. An anatomy‐driven shim region‐of‐interest selection allowed for improved B 0 ‐field homogeneity compared with a standard shim region‐of‐interest cuboid. Third‐order spherical harmonics terms were demonstrated to be beneficial for shimming of these myocardial B 0 ‐field inhomogeneities. Initial results from the in vivo implementation of a potential shim strategy were obtained. Simulated cardiac phase–specific shimming was performed, and a shim term‐by‐term analysis revealed periodic variations of required currents. Conclusion Challenges in state‐of‐the‐art B 0 shimming of the human heart at 7 T were described. Cardiac phase–specific shimming strategies were found to be superior to vendor‐supplied shimming.
No takes yet. Share an insight, caveat, or question.
Hock et al. (2020) studied Cardiac MRI at ultrahigh field (≥ 7T). Anatomy-driven shim region-of-interest selection and cardiac phase-specific shimming vs. Standard shim region-of-interest cuboid and vendor-supplied shimming was evaluated on B0-field homogeneity. Anatomy-driven shim region-of-interest selection and cardiac phase-specific shimming improved B0-field homogeneity compared with standard vendor-supplied shimming for human heart MRI at 7T.