This paper describes an experimental study which relates simultaneous elbow flexor-extensor electromyogram (EMG) amplitude to joint torque. Investigation was limited to the case of isometric, quasi-isotonic (slowly force-varying), nonfatiguing contractions. For each of the flexor and extensor muscle groups, the model relationship between muscle group torque contribution and EMG amplitude was constrained to be a sum of basis functions which had a linear dependence on a set of fit parameters. With these constraints, the problem of identifying the EMG-to-torque relationship was reduced to a linear least squares problem. Surface EMG's from elbow flexors and extensors, and joint torque were simultaneously recorded for nonfatiguing, quasi-isotonic, isometric contractions spanning 0-50% maximum voluntary contraction. Single-/multiple-channel unwhitened/whitened/adaptively whitened EMG amplitude processors were used to identify an EMG-to-torque relation, and then estimate joint torque based on this relation. Each unwhitened multiple-channel EMG-to-torque estimator had a standard error (SE) approximately 70% of its respective single-channel estimator. The adaptively whitened multiple-channel joint torque estimator had an SE approximately 90% of the unwhitened multiple-channel estimator, providing an estimation error approximately 3% of the combined flexion/extension torque range. The experimental studies demonstrated that higher fidelity EMG amplitude processing led to improved joint torque estimation.
No takes yet. Share an insight, caveat, or question.
Clancy et al. (1997) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: