The use of mass chromatogram peak centroids has been investigated as a means of deconvoluting the spectra of overlapping gas chromatography/mass spectrometry components. The peak centroids have been calculated with a precision of 0.04 scans (sd). This proved sufficient to allow deconvolution of the mass spectra belonging to two chemical components which were eluted 0.48 scans apart. For a more complex chromatography peak, it was possible to deconvolute the spectra of six components which were eluted within a 9 scan window. All the spectra produced by using this deconvolution mechanism agreed well with National Institute of Standards and Technology database spectra.