A compact sensing pixel for the determination of the localized complex permittivity at microwave frequencies is proposed. Implemented in the 40-nm CMOS, the architecture comprises a square patch, interfaced to the material-undertest sample, that provides permittivity-dependent admittance. The patch admittance is read out by embedding the patch in a double-balanced, RF-driven Wheatstone bridge. The bridge is cascaded by a linear, low-intermediate frequency switching downconversion mixer, and is driven by a square wave that allows simultaneous characterization of multiple harmonics, thus increasing measurement speed and extending the frequency range of operation. In order to allow complex permittivity measurement, a calibration procedure has been developed for the sensor. Measurement results of liquids show good agreement with theoretical values, and the measured relative permittivity resolution is better than 0.3 over a 0.1-10-GHz range. The proposed implementation features a measurement speed of 1 ms and occupies an active area of 0.15 × 0.3 mm 2 , allowing for future compact arrays of multiple sensors that facilitate 2-D dielectric imaging based on permittivity contrast.
No takes yet. Share an insight, caveat, or question.
Vlachogiannakis et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: