Assesses DKA risk trends in type 1 diabetes related to insulin pump use, indicating a decline over four decades.
Introduction: Continuous subcutaneous insulin infusion (CSII) in type 1 diabetes has been regarded as a major diabetic ketoacidosis (DKA) risk factor. We aimed to determine secular trends in risk since CSII implementation in the 1980s. Research Design and Methods: We assessed the relationship between time-varying CSII use and DKA events from 1983 to 2017 and by each decade in the 1441 Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications Study participants using crude and adjusted Cox proportional hazards models. Results: Time-varying CSII exposure was associated with significantly higher DKA risk in the 1980s (adjusted hazard ratio [HR] 5.81; 95% confidence interval [CI] 3.28–10.29; P < 0.001), but in the 2010s, this risk was not significantly elevated (adjusted HR 1.24; 95% CI 0.73–2.12; P = 0.43). Conclusions: DKA risk associated with CSII in type 1 diabetes has declined substantially since the 1980s such that the remaining risk in the past decade appears to be of low magnitude.