PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
August 1, 1999Review of Scientific Instruments213 citations

Ultralow-angle dynamic light scattering with a charge coupled device camera based multispeckle, multitau correlator

View Full Paper
LCLuca CipellettiDWDavid A. Weitz

Key Points

Key points are not available for this paper at this time.

Abstract

We use a charge coupled device (CCD) camera and a multi-tau software correlator to measure dynamic light scattering (DLS) at many angles simultaneously, from 0.07° to 5.1°. Real-time autocorrelation functions are calculated by averaging both over time and over CCD pixels, each corresponding to a different coherence area. In order to cover the wide spectrum of decay times associated with the large range of accessible angles, we adopt the multitau scheme, where the correlator channel spacing is quasilogarithmic rather than linear. A detailed analysis is presented of the effects of dark noise, stray light, and finite pixel area, and methods to correct the data for these effects are developed, making a CCD camera a viable alternative for a DLS detector. We test the apparatus on a dilute suspension of colloidal particles. Very good agreement is found between the particle radius derived from the CCD data, and that obtained with a conventional DLS setup.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Cipelletti et al. (1999) studied this question.

synapsesocial.com/papers/6a21e18965abd530f80d70f3https://doi.org/10.1063/1.1149894
Ask AI
Helpful
Bookmark
Share
View Full Paper