Key result
Adaptive ultrasonic measurement using custom-designed transmitted waveforms resulted in smaller uncertainty in blood vessel diameter and wall thickness measurements compared to conventional waveforms.
Why the study?
Does an adaptive ultrasonic measurement technique improve the signal-to-noise ratio and reduce measurement uncertainty compared to conventional techniques in measuring blood vessel diameter and wall thickness?
Does an adaptive ultrasonic measurement technique improve the signal-to-noise ratio and reduce measurement uncertainty compared to conventional techniques in measuring blood vessel diameter and wall thickness?
An adaptive ultrasonic technique significantly reduces measurement uncertainty for blood vessel diameter and wall thickness compared to conventional waveforms, potentially enabling examination of deep or small vessels.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating in animals; leaves open translation to human vascular ultrasound of deep or small vessels.
An adaptive ultrasonic technique for measuring blood vessel diameter and wall thickness is presented. This technique allows one to use a target-specific transmitted waveform/receiver filter to obtain a larger signal-to-noise ratio (SNR) in the received signal than conventional techniques. Generally, SNR of a received wave increases as the intensity of the transmit wave increases; however, because of the FDA limitations placed on the amount of transmit energy, it is important to be able to make the most efficient use of the energy that is available to obtain the best possible SNR in the received signal. Adaptive ultrasonic measurement makes the most efficient use of the energy that is available by placing the maximum amount of energy in the largest target scattering mode. This results in more energy backscatter from a given target, which leads to a higher SNR in the received waveform. Computer simulations of adaptive ultrasonic measurement of blood vessel diameter show that for a SNR of 0 dB in the transmitted waveform, the standard deviation of the diameter measurements for a custom-designed transmitted waveform is about two orders of magnitude less than the standard deviation of the diameter measurements using more conventional waveforms. Diameter and wall thickness measurement experiments were performed on a latex tube and a bovine blood vessel using both custom-made and conventionally used transmitted waveforms. Results show that the adaptively designed waveform gives a smaller uncertainty in the measurements. The adaptive ultrasonic blood vessel diameter and wall thickness measuring technique has potential applications in examining vessels which are either too deep inside the body or too small for conventional techniques to be used, because of the low SNR in the received signal.
No takes yet. Share an insight, caveat, or question.
Rafii et al. (1998) studied this question. Adaptive ultrasonic measurement technique vs. Conventional transmitted waveforms was evaluated on Uncertainty/standard deviation of diameter and wall thickness measurements. Adaptive ultrasonic measurement using custom-designed transmitted waveforms resulted in smaller uncertainty in blood vessel diameter and wall thickness measurements compared to conventional waveforms.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: