Key result
PVAC and nMARQ ablation systems for atrial fibrillation demonstrated comparable 1-year freedom from AF (79% vs. 80.7%), though nMARQ had shorter procedure times (94 vs. 81 min; P=0.0008).
Why the study?
Does nMARQ compared to PVAC improve procedural efficiency and clinical outcomes in patients undergoing PVI for symptomatic AF?
Observational (n=175)
Does nMARQ compared to PVAC improve procedural efficiency and clinical outcomes in patients undergoing PVI for symptomatic AF?
Both PVAC and nMARQ offer comparable acute and 1-year success rates for AF ablation, but nMARQ provides shorter procedure and burning times.
Comparable success rates with shorter nMARQ procedures; leaves open confirmation of efficiency gains in randomized AF ablation trials.
AIMS: We compare our experience with available circular multi-electrode catheters for atrial fibrillation (AF) ablation: PVAC(®), a phased radiofrequency system, and nMARQ™, an irrigated tip-CARTO-based technology. METHODS AND RESULTS: Prospective observational study of 175 consecutive patients with follow-up duration of at least 5 months who underwent pulmonary vein isolation (PVI) for symptomatic AF using PVAC(®) (n = 93, age 61.4 ± 9.8 years; 60% male, 13% persistent AF) vs. nMARQ™ (n = 82, age 63.2 ± 10.6 years; 67% male, 24% persistent AF). Procedure and radiation times were 94 ± 27 and 33 ± 13 min for PVAC(®) and 81 ± 18 and 30 ± 8.5 for nMARQ™ (P = 0.0008 and P = 0.18), respectively. The number of applications and the total burning time (min) were 20 ± 7 and 19 ± 6.7 for PVAC(®) and 16 ± 5.6 and 11 ± 4 for nMARQ™ (P < 0.0001 for both), respectively. In two nMARQ™ patients with small atria and pulmonary veins (PVs) and in two PVAC(®) patients with large PVs, the procedure failed; switching to the alternative technology was successful. Acute success rate was 97% for PVAC(®) and 95% for nMARQ™. There was one tamponade in nMARQ™ group and non-significant different minor complications for both techniques. One-year freedom from AF was 79 and 80.7% with PVAC(®) vs. nMARQ™, after one PVI, and 88 vs. 87.7% after two PVIs. CONCLUSION: Both technologies have short procedure and fluoroscopy times, comparable complication rates, and comparable acute and 1-year success rates. The number of applications and total procedure and burning times were shorter with nMARQ™. nMARQ™ was more suitable for larger atria and PVs. Thus, a patient-based pre-ablation anatomy definition is probably warranted for appropriate selection of technology type.
No takes yet. Share an insight, caveat, or question.
Laish‐Farkash et al. (2015) conducted an observational in Symptomatic atrial fibrillation (n=175). PVAC vs. nMARQ was evaluated on 1-year freedom from AF after one PVI. PVAC and nMARQ ablation systems for atrial fibrillation demonstrated comparable 1-year freedom from AF (79% vs. 80.7%), though nMARQ had shorter procedure times (94 vs. 81 min; P=0.0008).