Background: To determine whether using augmented reality with a head-mounted display (AR-HMD) would reduce deviations between planned and achieved reverse total shoulder arthroplasty (rTSA) glenoid baseplate inclination and version. Methods: Ten fresh frozen shoulders from 5 human cadavers, which were free from fractures or other bony pathologies were used. Computed tomography scans were acquired for each shoulder, and imported into image 3-dimensional processing software to plan rTSA, and notably to define the target inclination and version of the glenoid baseplate. Two experienced surgeons placed a 1.6-mm Kirschner wire on the glenoid baseplate insertion site in each shoulder (5 per surgeon) using conventional instruments, and the AR-HMD was used to measure the inclination, version, in addition to the number of outliers. Afterward, using the AR-HMD (Pixee Medical, Besançon, France) the surgeons drilled and inserted the Kirschner wire for the glenoid baseplate positioning, and computed tomography was used to measure the inclination, version, and number of outliers. Mean distance from entry point was -1.1 ± 1.7 mm in the superior-inferior axis, and 0.5 ± 0.9 mm in the anterior-posterior axis. Conclusion: AR-HMD significantly reduces the absolute error between achieved and planned inclination and version of the glenoid baseplate during rTSA, though further studies are required to confirm the benefits of this technology in clinical settings.