Answers are grounded in this paper and cite their sources.
A precision conductivity ratio measurement system (PCRMS) has been developed. Its use and principle of operation are presented and the principles, features, and specifications of the key components of the system are described in detail. Special emphasis is placed on the procedures for calibration and testing of the equipment. The overall results of evaluation show that the equipment has a sensitivity of 0.0001 S, a precision of better than <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">±0.0003</tex> S, and an accuracy of better than <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">±0.001</tex> S over the salinity range of 2-42.
No takes yet. Share an insight, caveat, or question.
Luo Jiuru (1985) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: