Key result
ActivPAL CREA validly measures walking cadence and ActiPASS tracks medium-fast walking versus manual counts.
Why the study?
Do the activPAL CREA v1.3 algorithm and ActiPASS 2023.12 software accurately detect steps compared to manual counts during a progressive treadmill-based laboratory protocol?
Do the activPAL CREA v1.3 algorithm and ActiPASS 2023.12 software accurately detect steps compared to manual counts during a progressive treadmill-based laboratory protocol?
The CREA v1.3 algorithm and ActiPASS 2023.12 software are valid tools for measuring step cadence during specific walking speeds.
Supports research use of activPAL CREA v1.3 for cadence; leaves open clinical translation and free-living validation.
Thigh-worn accelerometry is commonly implemented to measure step cadence. The default activPAL CREA algorithm is a valid measure of cadence during walking, but its validity during running is unknown. The ActiPASS software is designed to analyse tri-axial accelerometry data from various brands. We tested the validity of CREA v1.3 and ActiPASS 2023.12 to measure step cadence against manually-counted steps. Forty-five participants (26♀, 23.4 ± 4.6 years) completed 5 walking (6 min each, 2–4 mph) and 5 running treadmill (5–7 mph) stages (423 total stages completed). Based on equivalence testing, walking cadence (stages 1–5: 92–124 steps/min) from CREA was statistically equivalent (zone: <±2.2% of the manually-counted mean) to manual counts (92–125 steps/min). However, CREA underpredicted cadence during running stages (stages 6–10: 143–135 steps/min) by ~ 11–20 steps/min (p < 0.001). The ActiPASS-derived cadences were equivalent (zone: <±3.3%) to manual counts for all walking stages (99–127 steps/min) except Stage 1 (zone: ±10.5%). ActiPASS underpredicted cadences during running (stages 6–10: 137–153 steps/min) by ~ 10–16 steps/min (p < 0.001) compared to manual counts (stages 6–10: 154–164 steps/min). The CREA v1.3 algorithm is a valid measure of cadence during walking while ActiPASS 2023.12 is a valid measure of cadence during medium-fast walking. Further research is required to improve step cadence estimation across ambulation speeds.
No takes yet. Share an insight, caveat, or question.
Pellerine et al. (2024) studied this question. The activPAL CREA v1.3 algorithm validly measures walking cadence, while ActiPASS 2023.12 is valid for medium-fast walking compared to manual counts (154-164 steps/min).