We have demonstrated simultaneous correction for the optical interference of skin and fat in tissue spectra by using a two-distance fiber-optic probe. We obtained the correction by orthogonalizing the spectra collected at a long source-detector distance (SD) to the spectra collected at a short SD and mapped to the long SD space. The method was validated in tissuelike three-layer phantoms as well as preliminarily in human tissue. After the correction, a partial-least-squares model of the phantoms showed enhanced prediction performance.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2005) studied this question.