This paper introduces two techniques for controlling an articulated mechatronic laparoscope through the eyes of the surgeon during minimally invasive surgery. The system consists of a 2D eye tracking unit interfaced with a mechatronic laparoscope that has five controllable degrees-of-freedom (DoF) located at the distal end of a rigid shaft. Through the use of image feedback from a tip mounted camera, a closed-loop gaze contingent framework featuring two separate control techniques (“Individual Joint Selection” and “Automatic Joint Selection”) was developed. Under this framework, the location of a surgeon's 2D fixation point is converted into commands that servo the laparoscope. Experimental results illustrate the ability of both techniques to perform real-time gaze contingent laparoscope control. A key advantage of the proposed system is the ability to provide the operator with sufficient distal dexterity to achieve stable off-axis visualisation in an intuitive, hands-free manner, thus allowing other handheld instruments to be controlled simultaneously. Potential applications include Single Incision Laparoscopic Surgery (SILS) or Natural Orifice Trans-Endoluminal Surgery (NOTES), where the use of multiple instruments passing through a single incision presents both visualization and ergonomic challenges.