Widespread use of the gasoline additive methyl tert-butyl ether (MTBE) and the subsequent human exposure that follows have led to the need to quantify MTBE in a variety of complex biological matrixes. In this work, we demonstrate our latest MTBE quantification assay for whole blood and uncover previously unidentified contamination sources that prevented routine quantification in the low picogram per milliliter (parts per trillion, ppt) range despite a sensitive and selective analytical approach. The most significant and unexpected sources of contamination were found in reagents and laboratory materials most relevant to sample preparation and quantification. In particular, significant levels of MTBE were identified in sample vial septa that use poly(dimethylsiloxane) (PDMS)-based polymers synthesized with peroxide curing agents having tert-butyl side groups. We propose that MTBE is one of the byproducts of these curing agents, which cross-link PDMS via the methyl side groups. Residual MTBE levels of approximately 20 microg/septa are seen in septa whose formulations use these curing agents. Fortunately, these levels can be significantly reduced (i.e., <0.2 ng/septa) by additional processing. Performance achieved with this sample preparation approach is demonstrated using a mass spectrometry-based method to quantify blood MTBE levels in the low-ppt range.
No takes yet. Share an insight, caveat, or question.
Chambers et al. (2005) studied this question.